Guidance & Manufacturer's Declaration
Table 202 – Guidance and Manufacturer's Declaration
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The iPac is suitable for use in electromagnetic environment specified below. The customer or
user of the iPac should ensure that it is used in such an environment.
Immunity
Test
ESD
IEC 61000-4-2
EFT
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
Dips/Dropout
IEC 61000-4-11
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
16040-101 Rev. J
Electromagnetic Immunity
All Equipment and Systems
IEC 60601
Compliance
Test Level
±6kV Contact
±6kV Contact
±8kV Air
±8kV Air
±2kV Mains
±2kV Mains
±1kV I/Os
±1kV I/Os
±1kV Differential
±1kV Differential
±2kV Common
±2kV Common
>95% Dip
>95% Dip
for 0.5 Cycle
for 0.5 Cycle
60% Dip
60% Dip
for 5 Cycles
for 5 Cycles
30% Dip
30% Dip
for 25 Cycles
for 25 Cycles
>95% Dip
>95% Dip
for 5 Seconds
for 5 Seconds
3A/m
3A/m
Level
Environment
Floors should be wood, concrete or ceramic
tile. If floors are synthetic, the R/H should be
at least 30%.
Mains power quality should be that of a
typical residential, commercial or hospital
environment.
Mains power quality should be that of a
typical residential, commercial or hospital
environment.
Mains power quality should be that of a
typical residential, commercial or hospital
environment. If the user of the iPac requires
continued operation during power mains
interruptions, it is recommended that the iPac
be powered from an uninterruptible power
supply or battery.
Power frequency magnetic fields should be that
of a typical residential, commercial or hospital
environment.
(continued)
Electromagnetic
- Guidance
35