Perform the
temperature of the sample block(s) can exceed 100 ° C. If the instrument has been used
Calibration
recently, keep your hands away until the sample block(s) reaches room temperature.
1
4
3
2
5
6
Notes
™
Applied Biosystems StepOne
Installation, Networking, and Maintenance Guide
PHYSICAL INJURY HAZARD. During instrument operation, the
1. In the main menu of the StepOne
Instrument Maintenance Manager.
2. In the Instrument Maintenance Manager, click Dye in the navigation pane.
3. In the Dye screen, select System Dye Calibration.
4. Click Start Calibration.
5. In the Setup screen, select Plate 1 is loaded into the instrument.
6. Click Next.
7. In the Run screen, click Start Run
™
and StepOnePlus
Real-Time PCR Systems
Chapter 6 Maintain the System
Regular Maintenance
™
software, select Instrument
.
105