O P E R A T I N G I N S T R U C T I O N S
GM700 Ex
Laser Gas Analyzer,
Version with Measuring Probe

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GM700 Ex and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SICK GM700 Ex

  • Page 1 O P E R A T I N G I N S T R U C T I O N S GM700 Ex Laser Gas Analyzer, Version with Measuring Probe...
  • Page 2 Original document This document is an original document of SICK AG. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 3: Table Of Contents

    Information on installation in potentially explosive atmospheres..Tools required.................... 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 4 Operating states..............7.11.2 Select the ambient temperature range........O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 5 10.4.1 Communication fault between evaluation unit and receiver 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 6 13.4 Possible ambient temperatures............... 13.4.1 Ambient temperature ranges..........O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 7: About This Document

    8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 8: Symbols And Document Conventions

    Hazard which could result in property damage. Note Hints O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 9: Information Symbols

    Important technical information for this product Important information on electric or electronic functions 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 10: Safety Information

    – IEC 60079-19: Equipment repair, overhaul and reclamation O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 11 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 12 The person setting the system up is responsible for the safety of the system in which the device is integrated. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 13: Warning Information On The Device

    2 Type plate GM700 Ex Figure 2: Laser symbol on device 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 14: Intended Use

    Please con‐ tact SICK Customer Service. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 15 Must be kept for reference. Must be passed on to new owners. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 16: Product Description

    Min. gas temperature Max. gas temperature QR code Month/year O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 17: Device Variants

    Digital purge medium valve, 230 V, G3/8", 2 mm, Ex-p 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 18: Function Of The Pressurized Enclosure

    Further information on the purge air unit, see Operating Instructions of the purge air unit. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 19: Check Cycle

    The warning message remains until the next check cycle output is within the tolerance. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 20: Explosion Protection In Accordance With Atex

    After switching off the main power supply: Wait 20 minutes before opening the enclosure. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 21: Zone Of Use Gm700 Ex 3G Components

    Feed, optics purge air å Measuring channel Measuring channel, Ex-free æ 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 22: Pressurized Enclosure System

    GMM700 and GMA700 are secured additionally with sprocket discs. These sprocket discs may not be removed. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 23: Gmp Measuring Probe In Detail

    Further characteristics of the GMP measuring probe, see "Measuring probe", page 91 page 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 24: Transport And Storage

    Remove the protective stickers Set the locking device to the "close" position O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 25: Storage

    Store all components of the measuring system in a dry, clean room. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 26: Mounting

    • Local work safety regulations must be observed O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 27: Tools Required

    Preparing the sampling point • Feed and drain for protective gas 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 28: Gm700 Ex Scope Of Delivery

    Duct wall - SR-unit: depending on flange with tube length + measuring probe length O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 29: Overview Of Installation Steps (Duct-Side Preparation)

    8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 30 If necessary, attach duct insulation to protect the device from heat. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 31: Install The Evaluation Unit

    Ensure the protective gas inlet and outlet are in a horizontal axis. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 32 The pressure losses are noticeable when the purge medium flow rate is too low so that the device is possibly not completely pre-purged. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 33: Installing The Purge Air Unit

    For information on installing the purge air unit, see the Operating Instructions of the purge air unit. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 34: Electrical Installation

    This is also valid for plug connections. Disconnect the GM32 Ex and any connected devices from the voltage supply. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 35 Ensure the potential equalization is connected during all work on the device described in these Operating Instructions. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 36: Connection Overview

    Figure 13: GM700 Ex 3G / zone 2 connection diagram Power supply Warning signal, Ex-p control FS840 O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 37: Lines

    M16 x 1.5 Thread length 6 mm Height 23 mm 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 38: Connecting Interfaces

    Connections of pressure sensor: Pins 4 and 5 Pressure monitor for purge air monitoring: Pins 6 and 9 O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 39 Max. outer capacity Max. outer inductivity L/R ratio (inductivity/resistance) 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 40: Electrical Connection Of The Evaluation Unit Evu

    2 = digital output (NO contact) for maintenance request 3 = digital output (NO contact) for function check O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 41: Connecting The Connecting Hose To The Sr-Unit On The Evu

    Mark the disconnecting device as disconnecting device for the GM700 Ex. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 42: Connecting The Fs840 Ex Overpressure Control

    Check the gland screw fittings, sealing plugs and flanges for leak tightness and tight seat. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 43 92 and FS840 Operating Instructions. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 44: Electrical Connection Of Sr-Unit

    A potential equalization not connected properly or missing can lead to malfunctions in device operation. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 45: Connecting The Potential Equalization On The Sr-Unit

    Screw the grounding conductor tight on the screw terminal. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 46: Potential Equalization - Evaluation Unit And Ex Control Unit

    Potential equalization connection on Ex control unit Potential equalization connection on evaluation unit O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 47: Commissioning

    Required Operating Instructions: • Operating Instructions, measuring system GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 48: Material Required

    "Connecting the protective gas on the SR-unit", page 52 O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 49: Zero Adjust

    The flange fixture can be removed before installing the SR-unit. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 50 Sealing ring 10 cup springs Spherical washer Self-locking nut O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 51: Aligning The Measuring Probe In Flow Direction

    The measuring probe can be adjusted by turning the device flange. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 52: Feeding Protective Gas

    Thread G3/8" or • hose connection for 8 mm diameter O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 53 "Technical data for Pro‐ tective gas", page 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 54: Switch The Pressurized Enclosure On

    Look through visor (1) on the right enclosure side diagonally from above to the mir‐ ror with the target. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 55: Installing The Weatherproof Cover

    Screw the bottom mounting ring to the top mounting ring. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 56: Starting Measuring Operation

    "Menu tree Maintenance", page Set Mode to“Off”. Press “meas”. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 57: Operating States

    "Possible ambient temperatures", page 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 58: Operation

    To operate the measuring system via the operating elements: Ensure an Ex-free atmosphere exists when opening the GMA700. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 59: Function Buttons

    Function Back Returns to next higher level menu. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 60: Menu Tree Of The Evaluation Unit

    O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 61: Menu Tree Parameter

    Range high: 0 ... 999999 Cycle Out: No, Yes 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 62 GMM700/DSP ■ XXXXXXX XXXX ■ Service Not defined O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 63: Menu Tree Calibration

    Outputs the complete parameter set via the Push Diag RS232 Service interface 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 64: Operating Using The Serial Interface

    Start measuring operation/switch to measuring operation Syntax: GM700<<GOTO MEAS O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 65: Serial Interface Commands - Diag

    Use a 2 mm precision screwdriver to adjust the display contrast setting. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 66 OPERATION Figure 31: Rotary potentiometer for display contrast O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 67: Maintenance

    Wait 20 minutes after switching off the main power supply before opening the enclosure. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 68 O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 69: Maintenance Plan

    1W = Weekly, 1M = Monthly, 3M = Every 3 months, 6M = Every 6 months, Y = Yearly 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 70: Expendable, Wearing And Spare Parts

    ATEX certification becomes void. Also observe the maintenance instructions in the Pressurized Enclosure Manual. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 71: Maintenance Work On The Fs840

    Never open the device when the room is too small. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 72: Visual Check Of The Sender/Receiver Unit And Evaluation Unit

    Check grounding conductors are free from corrosion. Check flanges and screw fittings for tight seat. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 73: Cleaning The Optical Interfaces

    Flap the target holder upwards and check the sender/receiver unit window under‐ neath for contamination. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 74: Etalon Tool

    • Training by SICK Figure 35: Etalon tool O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 75: Filter Box Measurement To Check The Measuring Channels (For Dry Gases)

    Fasten the reflector enclosure to the filter box with 4 screws. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 76 When the measured value has stabilized again, read off and note the value. Leave operating mode “Box measuring” on the EvU with the return button. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 77: Checking The Evaluation Unit

    Observe safety information, see "Safety information on commissioning", page 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 78: Cleaning The Purge Air Unit

    Put the purge air unit back into operation. . O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 79: Troubleshooting

    Disconnect the GM32 Ex and any connected devices from the voltage supply. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 80: Monitoring And Diagnostic System

    Warning and error messages are stored in the built-in message memory of the control unit. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 81: Device Not Functioning

    Exchange evaluation unit and/or electron‐ ics board module. Contact Service. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 82: Communication Fault Between Evaluation Unit And Receiver

    If not successful, contact Service. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 83 SIG: M HIGH Measuring channel signal too high. Contact Service. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 84: Warning Messages

    Warm-up phase after switching on. Cool device. Ambient temperature too high. O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 85: Repairing Inadequate Purge Air Supply

    Unsuitable materials Check project planning. Corrosion on flange. 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 86: Decommissioning

    Set the lever on the purge air fixture to the “Close” position". O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 87: Decommissioning Ex-Relevant Subassemblies

    Protection when working on the stack or hot or aggressive sample gases 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 88: Preparing The Device Ready For Shipping

    Clean optical surfaces, see "Cleaning the optical interfaces", page O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 89: Environmentally Compatible Disposal

    Electronics: Capacitors, rechargeable batteries, batteries • Display: Liquid of LC-Display 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 90: Technical Data

    Power supply Supply via evaluation unit: 24 V DC O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 91: Measuring Probe

    Frequency: 50 Hz • Power input: ≤ 75 VA 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 92: Technical Data For Pressurized Enclosure System

    Approx. 6 minutes 12.5.5 Pressurized enclosure system settings O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 93 2 mbar (0.03 psi) (relative) Signal pressure 1.5 mbar (0.022 psi) (relative) 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 94: Dimension Drawings: Sender/Receiver Unit

    (9.00) Figure 37: GM700 Ex 3G sender/receiver unit (all dimensions in mm) O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 95: Dimension Drawings: Open Measuring Probe (Gmp)

    2.000 2.128 1.489 1.239 All dimensions in mm 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 96: Dimension Drawing, Purge Air Unit

    (1.69) Figure 39: Dimensions, purge air unit (all dimensions in mm (inch)) O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 97: Dimension Drawings, Evaluation Unit With Pressurized Enclosure System Fs840

    Figure 41: Mounting flange DN125 (all specifications in mm) 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 98: Dimension Drawing, Weatherproof Cover, Sender/Receiver Unit

    525,8 Figure 42: Weatherproof cover for sender/receiver unit (all specifications in mm) O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK Subject to change without notice...
  • Page 99: Annex

    –10 ... 40 263 ... 313 14 ... 104 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 100 32 ... 122 X Attention: These ranges may not be used for Ex applications! O P E R A T I N G I N S T R U C T I O N S | GM700 Ex 8019532/100Q/V1-1/2018-08 | SICK...
  • Page 101 ANNEX 8019532/100Q/V1-1/2018-08 | SICK O P E R A T I N G I N S T R U C T I O N S | GM700 Ex Subject to change without notice...
  • Page 102 Phone +36 1 371 2680 Phone +386 591 78849 E-Mail ertekesites@sick.hu E-Mail office@sick.si India South Africa Phone +91-22-6119 8900 Phone +27 (0)11 472 3733 Further locations at www.sick.com E-Mail info@sick-india.com E-Mail info@sickautomation.co.za SICK AG | Waldkirch | Germany | www.sick.com...