European Certifications; Rest Of The World Certifications - Emerson Rosemount Ultrasonic 3107 Quick Start Manual

Level and flow
Hide thumbs Also See for Rosemount Ultrasonic 3107:
Table of Contents

Advertisement

Quick Start Guide
Canadian Standards Association (CSA)
I6 Certificate Number: 02 CSA 1352094 X
Ex ia IIC
Intrinsically Safe when installed with certified barrier meeting transmitter
Entity parameters:
Ui=30V, Ii=120mA, Pi=0.82W, Ci=5nF, Li=27uH
Temperature Code T4 at Ta = –40 to 60 °C or T6 at Ta = –40 to 55 °C.
Special conditions of use
1. Risk of electrostatic charge build up on plastic surfaces. Clean only with a damp
cloth.

European certifications

I1 ATEX Intrinsic Safety
Certificate number: Sira 09ATEX2299X
II 1G, Ex ia IIC Ga
T6 (Ta = –40 to 55 °C), T4 (Ta = –40 to 60 °C)
Ui = 30V, li = 120mA, Pi = 0.82W, Li = 27μH, Ci = 5nF
IP66, IP68

Rest of the world certifications

I7 IECEx Intrinsic Safety
Certificate No.: IECEx SIR 09.0124X
Ex ia IIC Ga
T6 (Ta = –40 to 55 °C), T4 (Ta = –40 to 60 °C)
Ui = 30V, li = 120mA, Pi = 0.82W, Li = 27μH, Ci = 5nF
IP66, IP68
ATEX and IECEx conditions for safe use (X)
Model numbers covered:
3107*****I1****, 3108*****I1****, 3107*****I7****, and 3108*****I7****
('*' indicates options in construction, function and materials).
The following instructions apply to equipment covered by certificate number Sira
09ATEX2299X:
1. The equipment may be used with flammable gases and vapors with apparatus
groups IIA, IIB, and IIC, and with temperature classes T1, T2, T3, T4, T5, and T6.
2. Installation of this equipment shall be carried out by suitably trained personnel, in
accordance with the applicable code of practice.
3. The equipment is not intended to be repaired by the user and is to be replaced by
an equivalent certified unit. Repairs should only be carried out by the
manufacturer or approved repairer.
4. If the equipment is likely to come into contact with aggressive substances, it is the
responsibility of the user to take suitable precautions that prevent it from being
adversely affected, thus ensuring that the type of protection is not compromised.
Aggressive Substances e.g. acidic liquids or gases that may attack metals or
solvents that may affect polymeric materials.
Suitable Precautions e.g. regular checks as part of routine inspections or
establishing from the material's data sheet that it is resistant to specific chemicals.
5. The apparatus electronics is only certified for use in ambient temperatures in the
range of –40 to 60 °C for T4 or –40 to 55 °C for T6. It should not be used outside
this range.
22
July 2016

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount ultrasonic 3108

Table of Contents