Alarm Text
SpO
T SENSOR MALF
2
TELEMETRY STANDBY
TRANSMITTER MALF
TRANSMITTER OFF
TELE WEAK SIGNAL
Technical Alarms (Inoperative Conditions)
Priority
Condition
Hard
Malfunction of the SpO
sensor/adapter cable
Soft
Information Center standby
mode timer is active, or
patient was not returned to
telemetry coverage area.
There is no data from bed.
Hard
Transceiver malfunction
Hard
RF Auto Shutoff after 10
minutes of all leads of and
no SpO
Soft
2
sensor connected.
2
Patient is at outer
range of the radio
coverage area.
Telemetry pack is
receiving a weak
signal with high data
loss from the AP.
Condition exists for
multiple devices in a
specific area
Technical Alarms (INOPs)
What to do
Replace sensor.
Cancelled when patient is
removed from Standby.
Replace and notify service
provider.
Reattach ECG leads to
patient.
Reattach SpO2 sensor.
Press the Check
button.
Return patient to the
coverage area.
If patient is in close
proximity to AP,
replace telemetry pack.
Contact service.
The AP covering the
specific area is
suspect. Contact
service.
Alarms
5-13