Philips IntelliVue TRx Instructions For Use Manual page 171

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Battery
The transceiver battery status is not available. When the Check button is
pressed, the battery gauge on the transceiver and at the Information Center
indicates full power, regardless of actual battery strength.
Defibrillation
In the event of patient defibrillation, it may take several seconds for the
ECG trace to reappear on the screen.
Telemetry Overview
If NBP is being monitored, NBP values will be displayed at the Overview
window.
Telemetry Functions at the Information Center & TeleMon
Transceiver Operation when Connected to TeleMon
9-17