Spo Enable/Disable At Information Center - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

SpO
2
Enable/
Disable at
Information
Center
SpO
Auto
2
ON at
Information
Center
You can enable/disable SpO
"Patient Configurable Settings in Telemetry Setup" on page 9-4.
When SpO
is enabled, the Patient Sector and Patient Window of the
2
Information Center display a "T" next to the SpO
"SpO
T 90%") to indicate that the measurement was made via telemetry.
2
When SpO
monitoring is disabled, setting the sample rate to Spot Check or Off
2
at the Information Center or at TeleMon will help you conserve the
transceiver's battery life. See "Patient Configurable Settings in Telemetry
Setup" on page 9-4.
The SpO
parameter is automatically enabled at the IntelliVue Information
2
Center if a manual SpO
Spot Check mode or if the SpO
transceiver is in Continuous SpO
When a patient is discharged and the transceiver is in Continuous mode, the
SpO
parameter is disabled. To reactivate the SpO
2
from the transceiver, remember to do one of the following when a patient is
discharged:
– remove the SpO
reinsert the cable
– if using TeleMon, reset the transceiver to Manual mode.
The SpO
parameter Auto ON feature only needs to be reactivated when
Note—
2
the transceiver is in Continuous mode at discharge.
SpO
can always be enabled/disabled at the IntelliVue Information
Note—
2
Center.
monitoring at the Information Center. See
2
measurement is initiated at the transceiver while in
2
sensor is inserted into the transceiver while the
2
mode.
2
cable from the transceiver, wait 15 seconds, then
2
or
Measuring SpO
numeric (for example,
2
parameter Auto ON feature
2
SpO
Monitoring
2
2
8-19