Pairing with a Direct Connection to the MP5 Patient Monitor

Pairing with a Direct Connection to the MP5 Patient Monitor
10-2
Pairing with IntelliVue Patient Monitors
When the transceiver, assigned to a patient sector at the Information Center, uses
an interface cable to connect to an MP5 Patient Monitor, the Information Center
automatically pairs (if configured to do so) the two devices. Pairing using this
method provides 4 ECG waves at the MP5, along with SpO
a pleth wave. Lead selection and alarm limits may be changed either at the
Information Center or the MP5 Patient Monitor.
To connect the transceiver to the monitor, perform the following steps:
Step
Action
1
Remove the protective cover from the transceiver's service port.
2
Insert the interface cable into the service port.
3
Attach the interface cable to the monitor using the port labeled with
the
icon.
Warning
Warning
When the transceiver is directly connected to the monitor, arrhythmia
relearning is initiated, and again when the transceiver is disconnected from
the monitor.
Controls on the Transceiver, (e.g. Telemetry button to generate a Nurse
Call alarm) will be inactive when directly connected to the monitor.
When transceivers and MP5 Patient Monitors are paired directly via
Important—
an interface cable, your ECG source must be connected before connecting or
disconnecting the interface cable to maintain synchronized alarm settings. The
interface cable must be disconnected prior to unpairing the devices at the
Information Center or the monitor.