Monitoring During Leads Off; Ecg Fallback - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Monitoring during Leads Off

ECG
Fallback
Multilead
Analysis
Single Lead
Analysis
During routine monitoring, the electrode indicators also notify you if one or
more leads are not functioning. When a "Leads Off" condition occurs, the
transceiver automatically illuminates the indicator corresponding to the missing
lead.
Should the lead in the reference lead placement position (standard mode
Note—
- RL; EASI mode - green) become disconnected, the ECG signal is lost entirely.
After 10 minutes of this leads off condition, without the SpO
the transceiver will enter an RF Auto Shutoff state (if enabled). To correct the
condition, re-attach the lead to the electrode.
ECG Fallback and Extended monitoring states are supported for the transceiver
when the primary and/or secondary leads are in a "Leads Off" INOP condition.
Both of these states are entered into after 10 seconds of "Leads Off" in an
attempt to maintain monitoring and arrhythmia analysis.
ECG Fallback occurs when the primary lead is in "Leads Off" for 10 seconds
and a secondary lead is available.
If there is a "Leads Off" technical alarm in the primary lead for > 10 seconds,
the active secondary lead becomes the primary lead. The arrhythmia algorithm
switches the leads on the display, but relearn does not occur. When the "Leads
Off" condition is corrected, the leads are switched back to their original state.
For single lead analysis, if there are two leads available, the secondary lead is
made the primary lead until the "Leads Off" condition is corrected. The
arrhythmia algorithm performs a relearn using the available lead.

Monitoring during Leads Off

sensor connected,
2
ECG Monitoring
6-27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Intellivue trx +

Table of Contents