Measurement Specifications; Ecg - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Measurement Specifications

ECG

Parameter
ECG channel
transmitted Leads
3 electrodes
5 electrodes
5 electrodes,
EASI
6 electrodes
Resolution
ECG Input
Input Impedance
Input Dynamic Range
DC Offset Range
CMRR
Bandwidth +/- 3 dB
Specification
Channel #1 = I, II, or III
Channel #1 = II
Channel #2 = III
Channel #3 = MCL
Channel #1 = Va-i
Channel #2 = Va-s
Channel #3 = Ve-s
Channel #1= II
Channel #2 = III
Channel #3 = Va
Channel #4 = Vb
5 μV
Differential, defibrillator protected against 360
joules discharge into a 100 ohm load
> 5 megohms (@ 10 Hz
+/- 9 mV
+/- 320 mV
> 90 dB @ 50, 60 Hz
0.05 to 40 Hz