Connecting the ECG Cable

Connecting the ECG Cable

6-22
ECG Monitoring
Caution
Do not insert a leadset in the transceiver during a self-test. The leadset should
be inserted either before the transceiver is powered up, or after the device has
established connection with the Information Center.
Step
Action
1
Match the arrow on the ECG cable with the arrow on the Lead
Insertion Guide according to the lead type you have chosen, and
insert the ECG cable into the transceiver.
Important—
the transceiver, otherwise the device may not detect the proper
leadset type. When correctly inserted, the colored line at the
base of the cable is not visible.
3-wire
2
colored
line
IntelliVue TRx
M4841A
Make sure that the cable is pushed completely into
EASI, 3 5.6
Match the arrow on the cable
with the left arrow (labeled
EASI, 3) on the Leadset In-
sertion Guide.
Leadset is keyed for
Note—
only one insertion position.