Physiologic Alarms

Physiologic Alarms

5-4
Alarms
duration time has expired from the transceiver (see following directions) or from
the Information Center.
Step
Action
1
Press the
Telemetry
indicator turns off.
The Alarm Suspend icon is turned off.
The message "Alarms Suspended" (or "Alarms Paused") is
removed from the Information Center display.
2
Alternately, you can unsuspend alarms at the Information Center.
See IntelliVue Information Center Instructions for Use or Online
Help for directions.
For instructions on how to resume alarms from TeleMon, see the
TeleMon Instructions for Use.
Physiologic alarms indicate a life-threatening situation or a less urgent situation
such as heart rate beyond limits. There are no physiologic alarm signals
generated by the transceiver.
All physiologic alarms are generated at the IntelliVue Information Center,
and all alarm signals must be acknowledged at the Information Center.
Arrhythmia alarm chaining and other aspects of alarm behavior, such as alarm
levels, setting alarm limits, customizing arrhythmia alarm settings on a per
patient basis, switching individual measurement alarms on/off, and reviewing
alarm messages, are described in IntelliVue Information Center Instructions for
Use.
There are two levels of arrhythmia analysis available at the Information Center:
Basic and Enhanced. Enhanced analysis includes Basic alarms.
and
buttons simultaneously until the
Check