ST/AR ST Segment

SpO

2
Warning
Warning
It is possible that pacemaker pulses will not be detected when the ECG
analog output of a defibrillator is plugged into a bedside monitor. This can
result in the arrhythmia algorithm's failure to detect pacemaker non-
capture or asystole.
Warning
Warning
This device provides ST level change information; the clinical significance
of the ST level change information should be determined by a physician.
Warning
Warning
Always confirm Information Center observations with clinical observation
of the patient before administering interventions.
Warning
Warning
Using a sensor during MR imaging can cause severe burns. To minimize
this risk, ensure that the cable is positioned so that no inductive loops are
formed. If the sensor does not appear to be operating properly, remove it
immediately from the patient.

ST/AR ST Segment

2-11
Product Safety