Testing Transceiver Functionality

Self Test

In Case of
Failure
There are two tests of IntelliVue Transceiver functionality:
Self Test - performed automatically each time the transceiver is turned on.
Status Check - initiated manually by the clinician.
Warning
Warning
Do not use the transceiver for patient monitoring if it fails the Power On
Self Test.
Caution
Do not insert a leadset during a self-test. The leadset can be inserted before or
after the self-test.
A self test of the transceiver functions is automatically performed each time that
the transceiver is turned on (that is, when batteries are inserted). This test should
be performed before ECG leads and/or an SpO
Self Test
Auditory
Status
Signal
Passed
beep
Failed
beep beep
If any portion of the self test fails, the transceiver will attempt to report the
failure to the monitoring system. In case of failure, use another transceiver, and
contact your service provider.

Testing Transceiver Functionality

sensor is attached.
2
Visual Indicators
All indicators illuminate for 3 seconds.
One or more indicators do not light up.
Basic Operation
4-11