Alarm Text
* IRREGULAR HR
* MISSED BEAT
* MULTIFORM PVCs
* NON-SUSTAIN VT
* PACER NOT CAPT
* PACER NOT PACE
* PAIR PVCs
* PAUSE
Physiologic (Patient) Alarms
Priority
Condition
Yellow
Consistently irregular rhythm (irregular
R-R intervals).
Yellow
No beat detected for 1.75 x average R-R
interval for Heart Rate greater than 120,
or no beat for 1 second with Heart Rate
greater than 120 (non-paced patient only).
Yellow
The occurrence of two differently shaped
Vs, each occurring at least twice within
the last 300 beats as well as each
occurring at least once within the last 60
beats.
Yellow
A run of Vs having a ventricular Heart
Rate greater than V-Tach limit but lasting
for less than the V-Tach Run limit.
Yellow
No QRS for 1.75 x the average R-R
interval with Pace Pulse (paced patient
only).
Yellow
No QRS and Pace Pulse for 1.75 x the
average R-R interval (paced patient only).
Yellow
Two consecutive PVCs between non-
PVCs.
Yellow
No QRS detected for x seconds. Choices
of >1.5 to 2.5 seconds.
Physiologic Alarms
Source
ST/AR
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
5-7
Alarms