Alternate Cleaning And Disinfection Methods - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Alternate
Cleaning
and
Disinfection
Methods
Step
Action
3
Test the transceiver:
a. Put fresh batteries in the transceiver (without a leadset
b. Attach a leadset to the ECG port, and attach an SpO
c. Disconnect the Right Arm lead for standard ECG or the
d. Reconnect the electrode.
4
a. Connect the transceiver to TeleMon and observe the
The transceiver may be cleaned and disinfected using Metrex Cavi-Wipes, Cavi-
Wipes XL, CaviCide, or PDI Sani-Cloth
the instructions supplied by the manufacturers of these products.
attached) and close the battery door
Result: All six lead lights should flash, and one light
should remain on.
sensor to the SpO
port. If an ECG simulator is avail-
2
able, attach the ECG leads to the simulator and the SpO
sensor to yourself. At TeleMon, set the SpO
rate to Continuous.
Result: An ECG trace and SpO
visible on the Information Center display. All transceiv-
er lights should be off.
"I" electrode for EASI.
Result: The RA LED or the "I" lead LED should turn
on, and a Leads Off INOP should appear on the display
at the Information Center.
ECG waveform and SpO
play.
Result: The ECG waveform and SpO
be displayed on the TeleMon screen.
®
HB pre-moistened towelettes. Follow
Maintenance, Cleaning & Troubleshooting
Cleaning and Sterilization
sample
2
information should be
2
numerics on the TeleMon dis-
2
numerics should
2
2
2
11-11