Preinstallation Requirements
Space Requirements
3
In addition to space for the instrument itself, allow a 7.7-cm (3-in.) clearance on each side of the
instrument and a 16-cm (6.25-in.) clearance behind the instrument for air circulation.
4
Position the centrifuge so that the air diverter nearly touches the wall behind the centrifuge
(see
Figure
a. Place the power cord to one side of the air diverter.
b. To avoid damaging the power cord when installing or moving the centrifuge, be sure to
move the cord out of the way before pushing the centrifuge toward the wall.
Relative humidity should not exceed 95% (noncondensing).
Figure A.1 Dimensions of the Avant J-E Centrifuge
60.0
152.4
6.25
16.0
A-2
A.1).
31.5
80.0
3.0
7.7
in.
cm
34.0
86.4
3.0
7.7
25.0
63.5
PN SJ-IM-8AB