Emerson Rosemount 1057 Instruction Manual page 33

Three-input intelligent
Hide thumbs Also See for Rosemount 1057:
Table of Contents

Advertisement

Instruction Manual
LIQ-MAN-1057
5.4.3
Procedure - Configure Outputs
Under the Program/Outputs menu, the adjacent screen appears to
allow configuration of the outputs. Follow the menu screens in
configure the outputs.
5.4.4
Procedure - Assigning Measurements the Low and
High Current Outputs
The adjacent screen appears when entering the Assign function under
Program/Output/Configure. These screens allow you to assign a
measurement, process value, or temperature input to each output.
Follow the menu screens in
the outputs.
5.4.5
Procedure - Ranging the Current Outputs
The
adjacent
Enter a value for 4mA and 20mA (or 0 mA and 20 mA) for each output.
Follow the menu screens in
Figure 5-2 Configuring and Ranging the Current Outputs
Section 5: Programming the Analyzer - Basics
Figure 5-2
screen
appears
under
Figure 5-2
to assign values to the outputs.
Section 5: Programming the Analyzer - Basics
Figure 5-2
to assign measurements to
Program/Output/Range.
April 2017
to
S1: 1.234µS/cm
123.4°C
S2: 12.34pH
123.4°C
S3:123.4µS/cm
123.4°C
OutputM Assign
S1 Measurement
S1 Temperature
S2 Measurement
S2 Temperature
S3 Measurement
S3 Temperature
S1: 1.234µS/cm
123.4°C
S2: 12.34pH
123.4°C
S3:123.4µS/cm
123.4°C
Output Range
OM SN 4mA: 0.000µS/cm
OM SN 20mA: 20.00µS/cm
OM SN 4mA: 00.00pH
OM SN 20mA: 14.00pH
23