Ph/Orp/Ise (Codes -22, -32, -42) - Emerson Rosemount 1057 Instruction Manual

Three-input intelligent
Hide thumbs Also See for Rosemount 1057:
Table of Contents

Advertisement

Instruction Manual
LIQ-MAN-1057
1.4
pH/ORP/ISE (Codes -22, -32, and -42)
For use with any standard pH or ORP sensor. Measurement choices are pH, ORP, Redox, ammonia,
fluoride or custom ISE. The automatic buffer recognition feature uses stored buffer values and their
temperature curves for the most common buffer standards available worldwide. The analyzer will
recognize the value of the buffer being measured and perform a self stabilization check on the sensor
before completing the calibration. Manual or automatic temperature compensation is menu
selectable. Change in pH due to process temperature can be compensated using a programmable
temperature coefficient. For more information concerning the use and operation of the pH or ORP
sensors, refer to the product data sheets.
Model 1057 can also derive an inferred pH value called pHCalc (calculated pH). pHCalc can be derived
and displayed when two contacting conductivity sensors are used as sensor 1 and sensor 2.
Performance Specifications - Analyzer (pH Input)
Measurement Range [pH]: 0 to 14 pH
Accuracy: ±0.01 pH
Diagnostics: Glass impedance, reference impedance
Temperature coefficient: ±0.002pH/ °C
Solution temperature correction: Pure water, dilute base and custom.
Buffer recognition: NIST, DIN 19266, JIS 8802, and BSI.
Input filter: Time constant 1 - 999 seconds, default 4 seconds.
Response time: 5 seconds to 100%
Sensor Compatibility: Model 1057 is also compatible with SMART pH sensors from Rosemount.
Choose from 14 SMART sensor models for a wide array of applications and process
conditions.
Performance Specifications - Analyzer (ORP Input)
Measurement Range [ORP]
Accuracy: ± 1 mV
Temperature coefficient: ±0.12mV / °C
Input filter: Time constant 1 - 999 seconds,
default 4 seconds.
Response time: 5 seconds to 100% of final reading
Recommended Sensors For ORP: All standard
ORP sensors.
Temperature Specifications:
Temperature range
Temperature Accuracy, Pt-100, 0-50 °C
Temperature Accuracy, Temp. > 50 °C
Section 1: Description and Specification
:
-1500 to +1500 mV
0-150 °C
± 0.5 °C
± 1 °C
Section 1: Description and Specification
Fourteen SMART sensor models include PERpH-X,
TUpH and general purpose pH sensor families.
April 2017
5

Advertisement

Table of Contents
loading

Table of Contents