Damping Measurements
The Product normally applies a software filter to dampen
measurements in all functions except continuity. The
specifications assume that damping is turned on. The
damping method is a running average of the last eight
measurements. Fluke recommends leaving damping on.
Turning damping off may be useful when measurement
response is more important than accuracy or noise
reduction. To turn damping off, push the More Choices
softkey twice, then push the Dampen softkey so that Off is
shown. Push Dampen again to turn damping back on. The
default state is On.
Note
If a measurement falls outside a random noise window,
a new average is started. If damping is turned off, or
until measurements are fully damped, the symbol is
displayed.
Shop for Fluke products online at:
Source Mode
The operating mode (for example, MEASURE, SOURCE) is
shown on the display. If the Product is not in SOURCE
mode, push until SOURCE is shown. The Product
must be in SOURCE mode to change any of the SOURCE
parameters.
Source Electrical Parameters
To select an electrical source function:
1.
Connect the test leads as shown in Figure 18,
depending on the source function.
2.
Push for current, for dc voltage, h for
frequency, or q for resistance.
3.
Record the necessary output value, then push . For
example, to source 5.5 V dc, push 5 . .