Page 2
LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling.
Table of Contents Title Page Introduction ........................1 How to Contact Fluke ..................... 1 Standard Equipment....................... 2 Safety Information ......................2 Hazard Location Information/Approvals ..............3 Special Conditions for Safe Use ................3 Symbols ........................4 Display and Buttons ....................... 5 Operation ........................
Page 4
700G Series Users Manual Set Sample Rate ....................8 Set Tare ....................... 8 Function Lock ....................... 8 Supervisory Mode ..................... 8 Available Pressure Ranges ..................9 How to Set a Custom Engineering Unit or Scale............9 Battery Life ........................9 Maintenance ........................
Page 5
List of Tables Table Title Page Symbols ..........................4 Display and Buttons ......................5 1.888.610.7664 www.calcert.com sales@calcert.com...
Page 6
700G Series Users Manual 1.888.610.7664 www.calcert.com sales@calcert.com...
Page 7
List of Figures Figure Title Page The Product ........................... 5 Change the Batteries ......................11 1.888.610.7664 www.calcert.com sales@calcert.com...
Page 8
700G Series Users Manual 1.888.610.7664 www.calcert.com sales@calcert.com...
Introduction The 700G Series Pressure Gauges (the Product) are high-accuracy digital pressure test gauges. Accurate to 0.05 % FS, the Products can be used as a calibration reference, or in any application where high-accuracy pressure measurement is required. The Product features user-configurable functions that include: •...
700G Series Users Manual • Standard Equipment Replace the batteries when the low battery indicator () shows to prevent The Product ships with: incorrect measurements. • Protective Cover • Do not use and disable the Product if it • is damaged. Three AA Batteries (installed) •...
These areas are also any damage that can cause a safety concern. If referred to as hazardous locations, see NFPA 70 Article necessary, return the Product for evaluation to Fluke. 500. Refer to the How to Contact Fluke section.
Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information. 1.888.610.7664 www.calcert.com...
Page 13
Pressure Gauge Display and Buttons Display and Buttons The Display and Buttons are shown in Figure 1. The Buttons are explained in Table 2. PRESSURE 700G31 10K PSIG GAUGE 10K PSI 700 BAR gsn001.eps Figure 1. The Product 1.888.610.7664 www.calcert.com sales@calcert.com...
700G Series Users Manual Table 2. Display and Buttons Item Function Push to turn the Product on. Push again to turn it off. Zeros the display. In Configure Mode, push the button to move forward through the menus. Note ...
Page 15
Pressure Gauge Table 2. Display and Buttons (cont.) Item Function Item Function ¼ inch NPT Connector MIN MAX records minimum and maximum Pressure Display pressure values and saves them in memory. Push to show maximum Engineering Units (MAX) indication.
700G Series Users Manual engineering units. See the Supervisory Mode section for Operation instructions to set up custom units. The subsequent sections tell you how to operate the Set Auto Off Product. Push to turn on the Product. Auto Off can be set in 1-minute increments from 1 to 30 The analog bar graph at the bottom of the display shows minutes or you can turn off the function for continuous the applied-pressure level relative to the full range of the...
Pressure Gauge Operation Set Sample Rate Function Lock This function finds how often pressure is sampled and the When set, access to each of the settable parameters display is updated. Selections are 0.5, 1, 3, and 10 above can be turned “off” to prevent unauthorized samples/second.
700G Series Users Manual each parameter. To move to the next parameter, push How to Set a Custom Engineering Unit or Scale . The last menu selection in Supervisory mode is SET FACTR. You can set a multiplier factor from 0.001 to You can access, lock, or unlock these functions: 100 to make a custom scale.
Pressure Gauge Maintenance Maintenance How to Change the Batteries How to Clean the Product Warning To prevent possible electrical shock, fire, or Clean the Product with a soft cloth dampened with water personal injury, batteries must only be or water and weak soap. changed in an area known to be non- Caution hazardous.
700G Series Users Manual Accessories RS-232 Interface The Product includes an RS-232 interface. Remove the Product holster and the input jack is on the back of the Product. You can use serial communication to configure and calibrate the Product and move measurement data from the Product to a PC.
Pressure Gauge Specifications Specifications Available Input Ranges See Ranges and Resolution table for available ranges in psi plus equivalent ranges and resolution for all engineering units. Accuracy 700G Ranges Positive Pressure ................±0.05 % FS Positive Pressure (700G01, 700G02) ..........±0.1 % FS Vacuum ...................
700G Series Users Manual Media Compatibility 700G01, 700G02, 700G04, 700G05, 700RG05 ......any clean dry non-corrosive gas All other ranges, 15 psi to 1000 psi..........any liquids or gases compatible with 316 stainless steel Above 1000 psi ................any non-flammable, non-toxic, non-explosive, non- oxidizing liquid or gas compatible with 316 stainless steel Environmental Operating Temperature ..............
Pressure Gauge Specifications Mechanical Specifications Dimensions ..................11.4 x 12.7 (cm), depth = 3.7 cm (4.5 x 5 (in), depth= 1.5 in) Pressure Connection ................¼ in NPT Male Housing .................. Cast ZNAL Display 5-1/2 Digits, 16.53 mm (0.65 in) high 20-Segment bar graph, 0 to 100 % Power Battery ...................