753/754
Users Manual
9.
Push  for MEASURE/SOURCE mode. The display
changes to the screen shown in Figure 28.
Figure 28. Process Instrument Calibration Screen
10. Push the As Found softkey, followed by the
Instrument selection ().
70
Shop for Fluke products online at:
The display changes to the screen shown in Figure 29.
Figure 29. Process Instrument Calibration Screen 2
gks42s.bmp
11. Record values for 0% and 100% of 4.0 mA and
20.0 mA, in that sequence. Set Tolerance to 0.5% of
span.
MyFlukeStore