Page 2
Fluke Corporation (Fluke) warrants this product to be free from defects in material and workmanship under normal use and service for the life of the product. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, batteries or to any product which, in Fluke’s opinion, has been misused, altered, neglected or damaged by accident or abnormal...
Page 3
Safety Information This meter complies with EN 61010-1:1993, ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No. 1010.1-92 Overvoltage Category II. Use the meter only as specified in the Users Manual, otherwise the protection provided by the meter may be impaired. A Warning identifies conditions and actions that pose hazards to the user; a Caution identifies conditions and actions that might damage the meter.
Page 4
• Do not operate the meter if it is disassembled. Always operate the meter with the case top and bottom properly assembled. Disassembly procedures and warnings are in the 78 Automotive Service Manual. Service procedures are for qualified personnel only.
Page 7
3-10. DC and AC Current Test ..................3-12 3-11. Functional Description of Power Supply Components ........... 3-19 3-12. Voltage Levels ......................3-19 4-1. Fluke 78 Final Assembly ..................4-5 4-2. A1 Main PCA ......................4-7 5-1. Signal Abbreviations ....................5-3...
Chapter 1 Introduction and Specifications Title Page 1-1. Introduction .................... 1-3 1-2. Organization of the Service Manual ............1-3 1-3. Conventions.................... 1-4 1-4. Specifications ..................1-4...
• Schematic diagrams A meter under warranty will be promptly repaired or replaced (at Fluke’s option) and returned at no charge. See the registration card for warranty terms. If the warranty has expired, the meter will be repaired and returned for a fixed fee. Contact the nearest Service Center for information and prices.
Chapter 5. 1-4. Specifications Specifications for the Fluke 78 are in Table 1-1. Accuracy is specified for a period of one year after calibration, at 18°C to 28°C (64°F to 82°F) with relative humidity to 95%. AC conversions are ac-coupled, average responding, and calibrated to the rms value of a sine wave input.
Page 15
Output: Pulse to trigger Fluke 78 Thermocouple Type: K (Chromel vs. Alumel) (Fluke 80PK1) Not suitable for immersion in liquid. Accuracy: ±1.1°C (2°F) between 0°C to 260°C (32°F-500°F). Typically within 1.1°C (2°F) of NBS tables for temperatures between 40°C (40°F) to 0°C (32°F).
Page 16
Service Manual Table 1-1. Specifications (cont) Function Range Resolution Accuracy Burden Voltage (Typical) AC Volts* 4.000 V 0.001 V ± (2.5%+2) (45 Hz to 1 kHz) 40.00 V 0.01 V ± (2.5%+2) 300.0 V 0.1 V ± (2.5%+2) 300 V ±...
Page 17
Introduction and Specifications Specifications Table 1-1 Specifications (cont) Frequency Counter Sensitivity and Trigger Level Input Range Minimum Sensitivity Approximate Trigger Level (Rms Sine Wave) (DC Volts Function) 1 Hz to 5 Hz 5 Hz to 20 kHz 400.0 mV dc 400 mV 4000 mV dc 400 mV...
Theory of Operation Introduction 2-1. Introduction Chapter 2 provides theory of operation for the Fluke 78 Automotive Meter. An overview of circuit operation is presented in the form of functional block descriptions. Circuit descriptions cover the major circuit functions in more detail.
Service Manual 2-3. Circuit Descriptions Each of the functional blocks in Figure 2-1 is discussed in more detail in the following paragraphs. Refer to the schematic diagrams in Chapter 5 for circuit details not provided in this chapter. 2-4. Input Overload Protection The zinput is protected from overload by a network consisting of metal-oxide varistor (RV1), three current-limiting resistors (R1, R2, and RT1) and spark gap E1.
Theory of Operation Circuit Descriptions R1+RT1 Ω 9.996M 4 VAC ACTIVE HIGH FILTER 1.1111M CONVERTER zr02f.eps Figure 2-2. 4 V Range Simplified Schematic 2-8. Ohms Function When the 400-ohm range is selected, internal switches connect the resistor Z1-2 (9.996 MJ) to resistor Z1-6 (1.0001 kJ). (See Figure 2-3, 400-Ohm Range Simplified Schematic.) Then through switch contacts S1 5, 6, and 9, these resistors form a reference resistor of 1 kJ.
Service Manual 2-9. Continuity The continuity function is the 400 J range of the ohms function with no active filtering of the input signal. A comparator is used to turn on the beeper when the input drops below about 50 J. 2-10.
Theory of Operation Circuit Descriptions 2-11. Diode Test Function Q4 provides the source current for the diode test function. The input is sensed through R2. R2 and R9 form a 10:1 divider for measuring the voltage in diode test. A single beep sounds when the input drops below about 0.77 V;...
Page 26
Service Manual BUFFER INTEGRATOR COMPARATOR REFER TO WAVEFORM INPUT – VOLTAGE – – COMPARE CONTROL VOLTAGE REFERENCE 131.072 kHz CLOCK SWITCH AUTO INTE- READ READ HOLD ZERO GRATE +INPUT –INPUT MICROCOMPUTER INTEGRATE RAMP READ ZERO HOLD RAMP BASELINE 16.7 AUTO ZERO 8.33 REPEAT SAMPLE #1...
Theory of Operation Circuit Descriptions 2-15. Frequency Measurements A voltage comparator is used for both signal detection in frequency mode and threshold detection in continuity mode. In frequency mode, digital pulses from the voltage comparator are routed to the counter. Pressing the range push button while in frequency mode causes a range change in the primary function (ac or dc volts) that may change the sensitivity.
LCD. Both digital readings and an analog bar-graph display are presented in conjunction with annunciators and decimal points. Refer to the Fluke 78 Users Manual for information about the display. Table 2-1. Typical Voltage Levels and Tolerance (Referenced to Common) VBT + 3.7 ±0.2...
3-1. Introduction Chapter 3 contains maintenance information for the Fluke 78 meter and includes performance tests, calibration, general maintenance procedures, and troubleshooting. For operator maintenance, refer to the Fluke 78 Users Manual.
Page 32
Voltage: 8 V peak DC offset voltage: 0-4 V Frequency: 10 Hz-20 kHz Frequency Accuracy: .002% 100 mV DC Voltage Range Fluke 20 Series, 70 Series or 80 Accuracy: 0.5% Series Resolution: 100 µ V Thermometer Accuracy: 0.2 °C @ Ambient...
Maintenance Operator Maintenance 3-6. Fuse Test Use the following procedure to test the meter’s internal fuse: 1. Turn the rotary selector switch to the J position. 2. Plug a test lead into the z input terminal, and touch the probe to the 10 A input terminal.
3-12. Setup 1. Allow the Fluke 78 to stabilize to room temperature 23°C +/- 5°C (73°F +/- 9°F). 2. Check the fuse and battery, and replace them if necessary. (Refer to the battery and fuse replacement procedures in this chapter.)
Service Manual 3-13. Display Test To test the display, hold the pushbutton down and turn the Fluke 78 on. All segments will remain on as long as the pushbutton is held down. Check whether all display segments come on as indicated in Figure 3-2.
485 to 515 V AC 3-16. Frequency Test 1. Set the Fluke 78 rotary switch to HzB and press the pushbutton for 2 seconds to put the meter in the ac-coupled frequency function. 2. Connect the Function Generator output to the z and COM input terminals of the meter.
2. Referring to Table 3-5, set the High Accuracy Function Generator for the output indicated in the steps. 3. Verify that the Fluke 78 display reading is within the limits shown for each function. Note For tests: waveform = Triangle and Frequency = 100 Hz.
34.62 to 35.38 MJ 3-20. Continuity Test 1. Set the Fluke 78 rotary switch to w and press the pushbutton for 2 seconds to put the meter in the continuity test function. 2. Referring to Table 3-8, apply inputs as indicated. Verify that the Fluke 78 display and beeper indicate as shown in the table.
Service Manual 3-21. Diode Test 1. Set the Fluke 78 rotary switch to G and press the pushbutton for 2 seconds to put the meter in the diode test function. 2. Connect the DC Voltage Calibrator output to the zand COM input terminals of the meter.
Page 41
8. Select the temperature function on the Fluke 78, °C or °F to match the 80T-150U. Allow the readings to settle. The flashing reading of the Fluke 78 should be within 3 °C or 5.5 °F of the 80T- 150/DMM reading after taking into account the reading difference noted in step 6.
Service Manual 3-24. RPM80 Inductive Pickup Test (Optional Accessory) To test the RPM-80 inductive pickup, a triangle wave from a Function Generator will simulate automobile spark plug signals on a loop of wire containing a 10-ohm resistor. The pickup will be clamped to the wire, and output voltage from the pickup will be monitored by an oscilloscope.
Page 43
Maintenance Performance Tests OSCILLOSCOPE FUNCTION GENERATOR Ω OUTPUT RESISTOR WIRE LOOP (14 AWG) INDUCTIVE PICKUP zr08f.eps Figure 3-4. Setup for RPM80 Inductive Pickup Test 3-15...
Page 44
Service Manual OSCILLOSCOPE DISPLAY OF TRIANGLE WAVE 3V P-P µ OSCILLOSCOPE DISPLAY OF INDUCTIVE PICKUP OUTPUT 5.7V 10 ms zr09f.eps Figure 3-5. Typical Waveforms for Inductive Pickup Test 3-16...
5 times. 4. Set the DC Voltage Calibrator for an output of +3.5 V dc. 5. The Fluke 78 should display 3.500 V dc +/- 0.001. If necessary, remove the four case screws and case top and adjust R26. R26 (ADJUST FOR 3.500V DC ±...
3-26. Troubleshooting The procedures provided in these paragraphs will help isolate problems with the meter. When troubleshooting the Fluke 78 Automotive Meter, follow the precautions listed on the "Static Awareness" sheet to prevent damage from static discharge. 3-27. Surface Mount Assemblies The Fluke 78 Automotive Meter incorporates surface-mount technology (SMT) on the printed circuit assembly (pca).
Maintenance Troubleshooting Table 3-11. Functional Description of Power Supply Components Component Function Power supply startup device. Q8 provides Q7 base startup current. Q8 is always off during meter operation. VR1 provides the system reference voltage. It is used for the a/d converter reference and as a reference for both power supply regulators.
Service Manual 2. If Vrefh is correct, measure the voltage at U1 pin 6 (PS1). If Vps1 is not equal to 0 V +/-0.15 V, check R34 and R35. If Vps1 is still at an incorrect voltage, U1 is bad. 3.
Chapter 4 List of Replaceable Parts Title Page 4-1. Introduction .................... 4-3 4-2. How to Obtain Parts ................4-3 4-3. Manual Status Information..............4-3 4-4. Service Centers..................4-4 4-5. Parts Lists ....................4-4...
Introduction 4-1. Introduction This chapter contains an illustrated list of replaceable parts for the Fluke 78 Automotive Meter. Parts are listed by assembly; alphabetized by reference designator. Each assembly is accompanied by an illustration showing the location of each part and its reference designator.
4-5. Parts Lists The following tables list the replaceable parts for the 78 Automotive Meter. Parts are listed by assembly; alphabetized by reference designator. Each assembly is accompanied by an illustration showing the location of each part and its reference designator. The parts lists give the following information: •...
Page 53
List of Replaceable Parts Parts Lists Manual Status Information Ref or Option No. Assembly Name Fluke Part No. Revision Level Main PCA 919709 Table 4-1. Fluke 78 Final Assembly Reference Description Fluke Notes Designator Stock Number *MAIN PCA BT 1...
Page 54
Service Manual MP28 FLUKE 78 T&B VIEW XA11C.EPS Figure 4-1. Fluke 78 Final Assembly...
Page 55
List of Replaceable Parts Parts Lists Table 4-2. A1 Main PCA Reference Description Fluke Note Designator Stock Number CAP, POLYES, 0.01 µF, ±10%, 1000 V 822361 CAP, CER, 3.3PF, ±0.25 PF, 1500 V, COJ 904636 C 3, 8 CAP, CER, 150PF, ±5%, 50 V, COG,0805...
Page 56
Service Manual FLUKE 78-4001 zr12f.eps Figure 4-2. A1 Main PCA...
Page 59
Schematic Diagrams Table 5-1. Signal Abbreviations Number Name Description REFI REFerence Input AGND Analog GrouND REFH REFerence High plus power supply Power Supply 2 Power Supply 1 Power Supply 0 BIAS BIAS input Digital Output Port 2, bit 1 Diode Test Source OSCI OSCillator Input OSCO...
Page 60
Service Manual Table 5-1. Signal Abbreviations (cont) Number Name Description TEST TEST Port 0, bit 0 ClocK 1 ClocK 2 Icd backplane 1 Icd backplane 2 Icd backplane 3 Icd backplane 4 PUD* Power Up Digital output, active low Temperature Test RPM function input Function bit 0 Function bit 1...