Dc Voltage Measurement; Ac Voltage Measurement - Fluke 753 User Manual

Documenting process calibrator
Hide thumbs Also See for 753:
Table of Contents

Advertisement

753/754
Users Manual

DC Voltage Measurement

Range
Resolution
±3.00000 V
0.00001 V
0.0001 V
±30.0000 V
±300.00 V
0.01 V
Input Impedance: >4 MΩ
Maximum Input Voltage: 300 V, IEC 61010 300V CAT II
Temperature coefficient: (0.001 % of reading + 0.0002 % of range) / °C (<18 °C or >28 °C)
Normal mode rejection: >100 dB at 50 or 60 Hz nominal

AC Voltage Measurement

Range
Resolution
40 Hz – 500 Hz
3.000 V
0.001 V
30.00 V
0.01 V
300.0 V
0.1 V
Input Impedance: >4 M Ω and <100 pF
Input Coupling: AC
Maximum Input Voltage: 300 V, IEC 61010 300V CAT II
Temperature coefficient: 5 % of specified accuracy / °C (<18 °C or >28 °C)
Specifications apply for 9 % to 100 % of voltage range.
108
Shop for Fluke products online at:
% of Reading + Floor
1 Year
0.02 % + 0.00005 V
0.02 % + 0.0005 V
0.05 % + 0.05 V
% of Reading + Floor
1 Year
0.5 % + 0.002 V
0.5 % + 0.02 V
0.5 % + 0.2 V
MyFlukeStore
www.
2 Year
0.03 % + 0.00005 V
0.03 % + 0.0005 V
0.07 % + 0.05 V
2 Year
1.0 % + 0.004 V
1.0 % + 0.04 V
1.0 % + 0.2 V
1.888.610.7664