Installation of the probe by utilizing the ball valve set
Removing the probe
• Hold firmly the probe to avoid it being suddenly expulsed when releasing the lock not. May sure
nut to bend and damage the probe cable.
• Loosen slowly the lock nut with a spanner (spanner width 24) only till the probe is pushed out
by the overpressure in the process. Do not completely loosen the lock nut, but only as much as
necessary for the probe to slide.
• After the probe has been pushed out of the process up to the stop, close the ball valve.
• Remove the probe from the ball valve.
Observe the correct positioning of the sealing element 1 before reinstalling the probe.
14
metal sealing ring
(as a standard contained
with the ball valve set)
1/2" ball valve
lock nut
adapter
body
inside diameter
≥ 13.1 mm
(0.55")
clamping
sleeve
sealing
element 2
sealing
element 1
Operating instructions for EE310 Humidity / Temperature Transmitter