E+E Elektronik EE310 Operating Manual page 9

Humidity/temperature transmitter
Hide thumbs Also See for EE310:
Table of Contents

Advertisement

power supply +
analogue output
Option E5
M16x1.5
Modbus RTU
Option E6
Modbus RTU power supply +
analogue output
Option E12
Modbus RTU
Option AM3
analogue
output
Modbus RTU
(only with order code J3)
Fig. 7
Plug options for EE310 (front connector view)
External diameter of the supply cable for option AM3 and power supply + analogue ouput (E4 /
E6 / E12): 10-12 mm
Maximum wire cross section for connecting cable: 1.5 mm²
The AM3 option includes a 1.25 A fuse on the 100-240 V side. This fuse may not be replaced by
the user, but only by the E+E Service. National regulations for installation shall be observed.
External diameter of the cable for Modbus RTU female plug: 4 - 6 mm
Maximal wire cross section for connecting cable: 0.5 mm²
Operating instructions for EE310 Humidity / Temperature Transmitter
V+
n.c.
RS485 A
(=D+)
n.c.
RS485 A
(=D+)
power supply +
analogue output
n.c.
RS485 A
(=D+)
n.c.
power supply
phase
(L1)
Power supply 100 - 240 V AC
.
(0.39-0.47")
5
1
2
OUT 1
GND RS485
(shielded)
RS485 B
3
4
(=D-)
5
2
1
n.c.
GND RS485
(shielded)
RS485 B
3
4
(=D-)
5
2
1
n.c.
Modbus RTU
GND RS485
(shielded)
RS485 B
3
4
(=D-)
5
2
1
n.c.
Modbus RTU
neutral wire
3
(N)
1
n.c.
(AWG 16)
(AWG 21)
GND Output
OUT 2
GND
4
3
5
1
2
OUT 1
V+
Power supply + analogue output
GND Output
OUT 2
n.c.
4
3
5
1
2
OUT 1
n.c.
Analogue output
.
.
(0.16 - 0.24")
.
9

Advertisement

Table of Contents
loading

Table of Contents