3.3.3
T5: Remote probe up to 180 °C
EE310-T5 are transmitters with remote probe.
For mounting template of polycarbonate enclosure see chapter 3.1 Mechanical installation /
Fig. 4 Dimensions and mounting template of polycarbonate enclosure in mm (inch).
T working range probe:
Mounting the remote sensing probe of EE310-T5
For mounting the probe into a separation wall use the stainless steel mounting flange (see
chapter 8 Replacement parts / Accessories). The immersion depth is adjustable. The probe shall
be horizontal.
In case of mounting into a separation wall, it is of paramount importance for accurate measure-
ment to avoid T gradients along the probe. In case of large T difference between the two sides
of the separation wall, it is highly recommended to insert the probe completely (up to the cable
outlet) into the wall. Should this not be possible, place a thermal isolation layer on the part of the
probe outside the wall (on the cable side).
The stainless steel mounting flange is not appropriate for pressure tight mounting. For pressure
tight requirements use EE310-T10.
For probe hanging onto its cable from the ceiling in applications where condensation is likely to
happen it is important to avoid condense water getting from the cable to the probe and into the
sensing head. For this use the drip water protection.
temperature gradient along
the probe shall be avoided
probe into separation wall
Fig. 10
Mounting the remote probe of EE310-T5
Operating instructions for EE310 Humidity / Temperature Transmitter
(356 °F)
-40...+180 °C
electronics:
-40...+60 °C
-20...+50 °C
(-40...176 °F)
(-40...140 °F)
(with display)
(-4...122 °F)
drip water
protection
bow for draining
condense water
sealing
85
3.3"
hanging probe
11