• A CD and/or DVD Read/Write drive (CD or DVD burner); it is not sufficient to have a
CD-ROM or DVD-ROM drive.
And
• Roxio Drag-to-Disc software (part of Roxio Easy CD and DVD Creator Version 6.0 or
higher) installed and properly configured on the importing PC.
For more information, including important details on configuration of the Roxio software,
see
Appendix (D) Data Transfer Using CDs and
folder
This step is subsequent to data transfer via export from a Stratus instrument or from
another review station PC. Export can occur via direct connection or a preexisting office
network. See the configuration instructions for each of these options below.
Importing or Retrieving from a Network Server
Note: To import from a network server, the PC must be connected to the network
server, and both the server and the PC must be properly configured to communicate.
To retrieve from a network server, you must be archiving to the server from the
Stratus OCT. Users are responsible for network setup and maintenance, including
installation and configuration of all necessary hardware and software. Carl Zeiss
Meditec Technical Support cannot troubleshoot or repair problems with network
connectivity. See Appendix
Notes on Retrieving
Note: Stratus OCT does not permit archiving data to CD or DVD. Therefore, you can
retrieve archive data only from DVD-RAM or from a network drive.
Note: You can retrieve Stratus OCT data from archives made with any Stratus OCT
Software Version down to A1.1.
Note: For instructions to retrieve, see
How to Import
To import data, follow the steps under

Configuration for Export via Preexisting Office Network

This section explains how to configure the Stratus OCT and a review station PC to
communicate via your preexisting office network (a local area network or LAN). To do this,
the Stratus OCT and the review station PC each must be connected to the office network by
a standard network patch cable. It is the user's responsibility to acquire and install the
necessary cables.
WARNING: When networking the Stratus OCT, use only network cables with
an unshielded RJ-45 connector. Use of a shielded network cable in the Stratus
OCT could result in electrical shock to the patient and/or examiner.
Stratus OCT User Manual
Configuration Guide
DVDs.Importing from a local
(A) Networking Guidelines
Retrieve Scan Data
Import Data
on page 9-10.
for more information.
on page 8-11.
PN 2660021134133 A