• Although this instrument is designed for continuous operation, it should be turned off
when not in use for an extended period.

Embedded Windows License

Each Stratus OCT instrument is issued with an embedded Windows license. The license
number is of interest only to CZM service personnel. The license label is affixed underneath
the front computer cover, and is inaccessible to users.

Product Compliance

Complies with 93/42/EEC Medical Device Directive.
Complies with US and Canadian medical electrical system safety requirements.

Product Safety

• IEC 60601-1
• UL 60601-1
• CSA C22.2 No. 601.1-M90
This instrument is classified as follows:
• Class I Equipment – Protection against electrical shock.
• Type B – Degree of protection against electric shock of applied part (chin and
forehead rests).
• Ordinary Equipment (IPX0)– Degree of protection against ingress of liquids (none).
• Continuous Operation – Mode of operation.
WARNING: To prevent electric shock, the instrument must be plugged into an
earth grounded outlet. Do not remove or disable the ground pin. Only an
authorized Carl Zeiss Meditec service representative may install the
instrument.
WARNING: Do not use the printer or the instrument with an extension cord or
a power strip (multiple portable socket outlet). For additional safety, do not
plug the printer and the instrument into the same wall outlet. Failure to
observe this warning could result in electrical shock to the patient and/or
examiner.
WARNING: Do not open the instrument covers. Opening the instrument
covers could expose you to electrical and optical hazards.
Stratus OCT User Manual
Introduction
1-11
PN 2660021134133 A