Navigator 600 Silica
Single-stream
Appendix B – Diagnostics and Alarms
B.1 Analyzer Diagnostic Information
The analyzer is programmed to display diagnostic messages to
provide information on servicing requirements and any other
conditions that develop during operation. The following tables
list icon types, diagnostic messages, relay status (on or off) and
any operations that are inhibited if the condition occurs.
Diagnostic icons conform to NAMUR 107.
The messages listed below are shown by icon type and then
alphabetically. All diagnostic messages displayed on the
analyzer are added to the analyzer's audit log.
Note. The 'Out of Service' relay has the following fail-safe
operation:
Normal operation – ON (energized)
Out-of-Service state – OFF (de-energized)
Icon
Monitor in Service
The analyzer is operating correctly.
20 Character User-defined Alarm Tag
User-defined alarm message.
Media Card Full
The SD card used for recording is full and archiving has stopped.
1. Download the data onto a PC.
2. Replace the SD card with an empty SD card as soon as
possible.
A / D Error (1, 2 or 3)
A hardware error exists on the main board.
Power the analyzer down, wait for 10 seconds and power back up.
Notes.
If the error persists contact your local ABB representative.
Replacement board (see Appendix H, page 88 – part number
AW600065).
Excessive Secondary Current in Electronics
Too much current is being drawn by the system electronics in; this
causes the analyzer to go into automatic shutdown.
Contact the Company or your local ABB representative.
Faulty Measurement Optics
Optical measurement out of limits.
1. Check the connection labelled 'MEASUREMENT CUVETTE'
on the Interconnection Board.
2. Replace optical measurement assembly (see Appendix H,
page 86 – part number AW600091).
Table B.1 Diagnostic Information, Relay Operations and Inhibits (Sheet 1 of 7)
66
Diagnostic message
Appendix B – Diagnostics and Alarms
Diagnostic Icon
NAMUR Status
Failure
Check function
Out of specification
Maintenance required
Out-of-
Cal. in
Hold
Service
Progress
Relay
Relay
Relay
ON
OFF
ON
OFF
ON
OFF
ON
Cal.
Failed
Inhibits
Relay
All calibrations
and cleans
All calibrations
and cleans
All calibrations
and cleans