5.6 Current Outputs

There are six tabs in the Current Outputs screen, one for each output:
Fields
Output Source
Output Range
Output Type
Out of Sample Ind.
Default Output
36
Description
The 'Output Source' field has several options:
None. No other fields are visible
Stream 1 – Additional fields appear
The high and low limits for the output range. Both values can be set independently (0 to 5000 ppb).
If the difference between the zero and span is too small, the output is very noisy.
The electrical high and low limits (0 to 22 mA).
For example, if the output range is set to 0 to 2000 ppb and the 'Output Type' to 4.00 to 20.00 mA,
at 0 ppb the output is 4.00 mA and at 2000 ppb the output is 20.00 mA.
Out of sample indicator. If set to 'Yes', the output goes to the default output value when an out of
sample condition occurs for the selected stream source.
The output value used when an out of sample condition occurs and 'Out of Sample Ind.' is set to
'Yes' (0 to 22 mA).