ABB Navigator 600 Silica User Manual page 52

Single-stream, colorimetric
Table of Contents

Advertisement

Navigator 600 Silica
Single-stream
Referring to Fig. 8.1:
A
1. Unlock catch
, use handle
section forward and lock it into place using catch
2. Release the pump tube clips on the downstream side to
prevent back-siphoning of reagents (and possible
contamination of solutions in the bottles) when platen
is unlatched.
3. Using a screwdriver, remove capstan screw
E
washer
,
4. Remove pump end plate
that the shaft-to-bearing adaptor is not lost.
5. Unlatch and rotate each platen
tubing. Slide the platens off the pivot pin.
6. Unlatch the other side of the pump tubing
the tubing.
7. Slide off and discard each capstan
8. Replace each capstan. Ensure the capstans are aligned to
maximize mixing effectiveness.
9. If not already fitted, clip cover
10. Clip the new tubing into place. Ensure the correct tubing is
used for each pump and the seals are in place:
Rear
Green
Blue
Front
Gray
Wet the mating surfaces with demineralized water and
squeeze between the two points shown below to ensure
that tubing clips into place correctly:
50
B
to slide the analytical
A
.
D
and
F
(pull off squarely), ensuring
C
, to access the pump
G
and discard
H
.
I
to the outer capstan.
Brown
Orange
Violet
Red
11. Ensure the tubing is aligned correctly over the rollers.
12. Refit the platens
Note. When fitting the pump end plate, ensure the
shaft / bearing adaptor engages with the central hole
in the pump shaft and that the pivot pin aligns with the
C
mating hole in the pump end plate
E
13. Refit washer
Tighten screw
14. Unlock the analytical section and slide it back into the
normal operation position. Lock the analytical section into
place.
For part numbers, refer to the Spares section – see Appendix H,
page 86.
Once the service has been completed, the annual service date
is reset and a purge routine is run automatically to purge the
system of air. After the purge sequence, perform a Baseline
calibration.
C
F
and pump end plate
F
.
D
and capstan screw
.
D
C
and clip each platen
IM/NAV6S/SS–EN Rev. E
8 Maintenance
.
into place:

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aw641

Table of Contents