Solution Replacement; Annual Service - ABB Navigator 600 Silica User Manual

Single-stream, colorimetric
Table of Contents

Advertisement

Navigator 600 Silica
Single-stream

8.1.2 Solution Replacement

The analyzer has predictive capability that notifies when the
solutions require changing. For details of the reagent, clean and
calibration solutions, refer to Appendix A, page 64.
The analyzer is taken offline when solution replacement is in
progress.
Note. To prevent air getting into the reagent lines, ensure
the peristaltic pumps are switched off prior to lifting the
reagents lines out of the reagent bottles. This is done either
automatically by entering the solution replacement page or
manually through the Manual Test Settings menu.
Fields
Description
Prompts for the reagents to be changed.
When all reagent solutions have been
changed, the analyzer purges the reagent
Replace All
lines so that the fresh reagent solutions enter
Reagent
the reaction coil. The analyzer then goes
Solutions
through a recovery sequence and then
prompts the user to instigate a calibration. A
Baseline calibration must be carried out after
replacing reagent solutions.
Replace
Prompts
Secondary
solution to be changed. Once changed the
Cal.
date of the replacement in the Diagnostic
Solution
Display is reset.
Prompts for the cleaning solution to be
Replace
changed. Once changed the date of the
Cleaning
replacement in the Diagnostic Display is
Solution
reset.
Opens the 'Calibration' window – see
Section 7.5, page 47 to enable the analyzer
Calibration
to be re-calibrated by running a Baseline
calibration routine.
IM/NAV6S/SS–EN Rev. E
for
the
secondary
calibration

8.1.3 Annual Service

The analyzer is taken offline and the pumps are stopped when
an annual service is in progress.
The annual service requires the pump tubing and all capstans to
be replaced. Once the service has been completed, the last
annual service date is reset and a purge routine is run
automatically. After each annual service, perform a manual clean
and, 6 hours later, a baseline calibration.
It is recommended that the pump tubing and pump capstans in
the annual pump refurbishment kit are replaced after one year of
operation.
Fig. 8.1 Pump Tubing and Capstan Replacement
8 Maintenance
49

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aw641

Table of Contents