Navigator 600 Silica
Single-stream
Icon
Solution Replacement
The analyzer is shut down while solution is replaced.
Scheduled Secondary Cal. Missed
The scheduled secondary calibration has been missed due to either
a manual operation being performed or an out-of-service condition.
The next calibration due date is updated automatically.
The next calibration is performed at the correct date and time
if the analyzer is in service.
Cleaning In Progress
A clean routine is being run.
A progress bar on the cleaning page indicates the time to
completion.
Calibration In Progress
A calibration routine is being run. A progress bar indicates the time
to completion.
1. Wait for the routine to be completed.
2. Cancel the routine and allow the analyzer to complete the
recovery routine.
In Annual Service
The analyzer has been taken off-line for the annual service.
Complete the servicing tasks and press the
to reset the annual service alarm and place the analyzer on-line
(after a purge and recovery period).
In Manual Test Setting
The analyzer has been taken off-line while manual test settings are
checked / adjusted.
1. Complete the manual test settings routine.
2. When completed, place the analyzer back on-line by exiting
via 'Normal' mode at the 'Exit via Test Mode / Normal Mode'
prompt.
Table B.1 Diagnostic Information, Relay Operations and Inhibits (Sheet 5 of 7)
70
Diagnostic message
key when finished
Appendix B – Diagnostics and Alarms
Out-of-
Cal. in
Hold
Service
Progress
Relay
Relay
Relay
OFF
ON
OFF
ON
ON
OFF
ON
OFF
ON
Cal.
Failed
Inhibits
Relay
Scheduled
calibrations
and cleans
Scheduled
calibrations
and cleans
Scheduled
calibrations
and cleans
Scheduled
calibrations
and cleans
Scheduled
calibrations
and cleans
IM/NAV6S/SS–EN Rev. E