Note: Depending on your system configuration, the inject start signal can be transmitted
through the IEEE-488 interface or the analog-out/event-in connectors on the 2410
refractometer rear panel. For information on non-IEEE-488 connections, see Section 3.3,
Making Non-IEEE-488 Signal Connections.
Table 3-2 summarizes the inject start connections for different system configurations.
Note: If multiple devices in your system require an inject start signal, connect trigger wires
from the same (inject out) terminal on the injector to each device.
Table 3-2 Waters 2410 Refractometer Inject Start Connections
Inject Start Output Source
Waters 715, 717, and 717plus,
and 2690 Separations Module, on
the IEEE-488 bus
3
Waters 715, 717, and 717plus not
on the IEEE-488 bus
Waters 2690 Separations Module
not on the IEEE-488 bus
Waters 712 Autosampler
Waters manual injector, or
third-party manual injector or
autosampler
3-6
Making Signal Connections
Inject Start Input Connection
(on the 2410 Refractometer)
IEEE-488 interface (see Section 3.2.1, Connecting to a
Waters Data System Using the IEEE-488 Bus)
Note: If you are using the Waters 845 or 860 Data
System, you must program the multi-method to Start
By LAC/E (refer to the ExpertEase Reference Guide
for details).
Chart Mark and Ground
Chart Mark and Ground or Auto Zero and Ground
Chart Mark and Ground
Chart Mark and Ground