Dc V (V) Performance Verification Tests - Keysight U1242C Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1242C:
Table of Contents

Advertisement

1
Calibration Procedures
Table 1-3

DC V (V) performance verification tests

Test function
DC mV (mV)
Turn the rotary switch to the
DC V (V)
Turn the rotary switch to the
[a] The impedance may be set to >1 GΩ from the Setup mode; the default value is 10 MΩ in parallel with 100 pF (nominal). The accuracy of DC
mV is specified after the Null function is used to subtract the thermal effect (by shorting the test leads) before measuring the signal.
[b] Before turning the rotary switch, you must enable the mV measurement function from the Setup menu first. For enabling the mV
measurement function from the Setup menu, refer to the U1241C/U1242C User's Guide.
26
position.
1
2
position.
V
5520A/5522A
Range
output
[a][b]
100 mV
100 mV
[a][b]
600 mV
600 mV
1000 mV
1000 mV
10 V
10 V
100 V
100 V
1000 V
1000 V
Keysight U1241C/U1242C Service Guide
Error from nominal 1 year
U1241C
U1242C
±0.11 V
±0.11 V
±0.7 V
±0.7 V
±1.1 V
±1.1 V
±0.011 V
±0.011 V
±0.11 V
±0.11 V
±1.1 V
±1.1 V