1
Calibration Procedures
Table 1-15
Adjustment input values (continued)
Test function
Capacitance
Diode
DC μ.mA
AC μ.mA
46
Step
Open
1000.0 nF
10.000 μF
100.00 μF
1000.0 μF
10.000 mF
Short
2.100 V
Open
1000.0 μA
10.000 mA
100.00 mA
1000.0 mA
1000.0 μA
1000.0 μA (55 Hz)
10.000 mA
10.000 mA (55 Hz)
100.00 mA
100.00 mA (55 Hz)
1000.0 mA
Reference value
oPEn
100.0 nF
1000.0 nF
1.000 μF
10.000 μF
10.00 μF
100.00 μF
100.0 μF
1000.0 μF
1.000 mF
10.000 mF
SHor
1.000 V
oPEn
1000.0 μA
10.000 mA
100.00 mA
300.00 mA
100.0 μA (55 Hz)
1.000 mA (55 Hz)
10.00 mA (55 Hz)
030.0 mA (55 Hz)
300.0 mA (55 Hz)
Keysight U1241C/U1242C Service Guide
Valid reference input
Open terminals
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
Short V Ω / COM terminals
0.9 to 1.1 x Reference value
Open terminals
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.7 to 1.3 x Reference value
0.9 to 1.1 x Reference value