Capacitance Performance Verification Tests; Resistance Performance Verification Tests - Keysight U1242C Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1242C:
Table of Contents

Advertisement

1
Calibration Procedures
Table 1-10

Capacitance performance verification tests

Test function
Capacitance
Turn the rotary switch to the
View
.
Shift
Esc
Table 1-11

Resistance performance verification tests

Test function
Resistance
Turn the rotary switch to the
[a] The accuracy of the 1000
32
position and press
position.
Ω
range is specified after Math Null, which is used to subtract the test lead resistance and the thermal effect.
5520A/5522A
Range
output
1000 nF
1000 nF
10 μF
10 μF
100 μF
100 μF
1000 μF
1000 μF
10 mF
10 mF
5520A/5522A
Range
output
[a]
1000 Ω
1000 Ω
10 kΩ
10 kΩ
100 kΩ
100 kΩ
1000 kΩ
1000 kΩ
10 MΩ
10 MΩ
25 MΩ
100 MΩ
100 MΩ
Keysight U1241C/U1242C Service Guide
Error from nominal 1 year
U1241C
U1242C
±10.5 nF
±10.5 nF
±0.105 μF
±0.105 μF
±1.05 μF
±1.05 μF
±12.5 μF
±12.5 μF
±0.125 mF
±0.125 mF
Error from nominal 1 year
U1241C
U1242C
±2.2 Ω
±2.2 Ω
±0.022 kΩ
±0.022 kΩ
±0.22 kΩ
±0.22 kΩ
±2.2 kΩ
±2.2 kΩ
±0.082 MΩ
±0.082 MΩ
±0.41 MΩ
±0.41 MΩ
±3.03 MΩ
±3.03 MΩ

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1241c

Table of Contents