Calibration Procedures
Valid adjustment input values
Adjustment can be accomplished using the following input values.
Table 1-15
Adjustment input values
Test function
AC V
[a]
Vsense
44
Step
1000.0 mV (55 Hz)
1000.0 mV (1 kHz)
1000.0 mV
1000.0 mV (55 Hz)
1000.0 mV (10 kHz)
10.000 V
10.000 V (10 kHz)
100.00 V
100.00 V (10 kHz)
1000.0 V
1000.0 V (10 kHz)
Hi.10 V
Lo.30 V
Reference value
100.0 mV (55 Hz)
10.000 V (55 Hz)
10.000 V (1 kHz)
1.000 V (55 Hz)
10.000 V (55 Hz)
10.00 V (55 Hz)
100.00 V (55 Hz)
100.00 V (1 kHz)
100.0 V (55 Hz)
1000.0 V (55 Hz)
1000.0 V (1 kHz)
10 V (55 Hz)
30 V (55 Hz)
Keysight U1241C/U1242C Service Guide
Valid reference input
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
0.9 to 1.1 x Reference value
High sense
Low sense