Performance Verification Tests - Keysight U1242C Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1242C:
Table of Contents

Advertisement

Performance Verification Tests

Use the performance verification tests to verify the measurement performance of
the multimeter. The performance verification tests use the multimeter's
specifications listed in the U1241C/U1242C datasheet
(http://literature.cdn.keysight.com/litweb/pdf/5992-0848EN.pdf).
The performance verification tests are recommended as acceptance tests when
you first receive the multimeter. The acceptance test results should be compared
against the 1-year test limits. After acceptance, you should repeat the
performance verification tests at every calibration interval.
If the multimeter fails the performance verification tests, adjustment or repair is
required.
Ensure that you have read the
NOTE
the performance verification tests.
Table 1-2
Z
LOW
Test function
[a]
Z
(DC V)
LOW
Turn the rotary switch to the
View
.
Shift
Esc
Z
(Input impedance)
LOW
Turn the rotary switch to the
[a] Functional test only.
Keysight U1241C/U1242C Service Guide
(DC V) performance verification tests
Z
LOW
position and press
[a]
Z
LOW
position.
"Test Considerations"
5520A/5522A
Range
output
1000 V
30 V
1000 V
1.8 kΩ
Calibration Procedures
on page 24 before running
Error from nominal 1 year
U1241C
N/A
N/A
±0.180 kΩ
1
U1242C
±0.7 V
25