Test Procedure

3.Test for glucose
Firmly insert the glucose test glucose test strip into the test port
of Element Multi meter and then the power will automatically
turn on with the code number, temperature and time.
Make sure code number on display of the meter correspond
with that on the tset strip vial. The color coding label of the back
on the test strip automatically indicates the code number in this
system.
The symbol of blood is blinking.
Once the sample is applied, the meter displays
3, 2, 1 in order.
Then, the meter completes the test with the test result.
Caution
1. If the code is not displayed after 3 seconds, pull the test strip out of the port and restart
the procedure from the beginning.
2. If the codes on the display and that on the test vial do not match, try another new strip.
If the mismatch persists, please contact your local representative for help.
3. If the meter does not power on, pull the test strip out of the port and re-insert the test strip.
4. Avoid testing under direct sunlight for more accurate test result.
5. If you apply your sample too early, error message will appear on the display(refer to
page 47)
• 38 •

Glucose test strip Discard Function

1. After checking your test result, turn the meter upside down and slide the
ejector button twice forward to remove the test strip from the meter.
2. Discard the used strip and lancet to a proper place, according to the local
regulation.
Caution
1. If you push the discard button forward too much, it may be damaged.
2. Do not give strong impact to meter.