Pos: 35.9 /Überschriften/1.1.1/1-spaltig/Neuabgleich der Spulentemperatur @ 16\mod_1196859777171_3101.doc @ 144585
7.4.3

Readjusting the coil temperature

Pos: 35.10 /Konfiguration, Parametrierung/Durchfluss/FSM4000/Erweiterte Diagnosefunktionen/Neuabgleich Spulentemperatur @ 16\mod_1196859641593_3101.doc @ 144562
If a readjustment is required, make sure that the flowmeter sensor and coil are relatively close
to ambient temperature. You might need to allow the flowmeter to cool off overnight.
Work steps:
Switch on the transmitter.
Under "Diagnostics / Adjustment / CabLength", enter or check the cable length.
Enter the ambient temperature in the submenu "Diagnostics / Adjustment / Coil Temp".
Store the new reference resistance in the submenu "Diagnostics / Adjustment / Transfer
RT20".
Pos: 36 /Konfiguration, Parametrierung/Durchfluss/FSM4000/Software-Historie/Software-Historie @ 16\mod_1196860571562_3101.doc @ 144633
7.5

Software history

7.5.1

For transmitters without communication or HART protocol

Software Version Type of changes
D184B140U02
B.10
OEM software
B.11
Improve min. contrast limit.
After changing from 50 → 60 Hz the
system data now displays the correct
frequency.
B.12
Shortened Finnish texts.
Aut. simulator detection for counter
management improved.
B.14
Improved FRAM management.
B.20
Added Turkish as new language.
Added error E (DC too high)
Updated flowmeter sensor 10D1462/72.
B.22
Always display the flowrate with 4 decimal
places.
B.30
Menu item for noise reduction
Menu item Meter factor
Support for flowmeter sensor
SE21, DN 1 ... DN 2
B.31
Hardware compatibility for C-level
hardware
B.32
Improved totalizer management
B.33/B.34
Improved FRAM management
C.10
Extended diagnostic functions.
Extended noise reduction by 2 more levels.
Software supports external zero point
adjustment via contact input.
Software D200S021U01
FSM4000
Parameterization
Documentation/supplements
Added documentation for the
additional points.
Added documentation for the
additional points.
Added documentation for the
additional points.