Supplementary information: Operating S4 with an older sensor
Warning – Electrical voltage risk!
When the housing is open, EMC protection is impaired and protection against contact is
suspended.
Power to all connecting cables must be switched off.
Adapting the transmitter to the flowmeter sensor 10D1422 (DN 3 ... 1000),
10DI1425 (≥ DN 500):
1. Switch off supply power. After waiting 40 minutes, remove the housing cover from the field
housing unit.
2. Replace the protective cover (1) by removing the 3 screws. Close switch S903 (3).
Fig. 58
1 Protective cover
2 Screws for cover