ABB FSM4000 Operating Instruction page 78

Electromagnetic flowmeter
Hide thumbs Also See for FSM4000:
Table of Contents

Advertisement

Parameterization
Submenu/parameter
Warning Mask
Software Rev.
Communication: 1
Submenu
Function Test
Memory Test
Pulse Output
0 Hz
Current Output
4 mA
Contact input
Off
Contact output
Off
Display
Test Mode
Off
78
Value range
Bitstring 0 OFF
020000100
Warning: 10
Reverse Q
Display only
Numeric entry
0 - 20 mA, 4 - 20 mA
0 - 10 mA, 2 - 10 mA
Off
On
Off
On
Off
On
Comment
This menu shows the screen for the warning register (TB
Index 96).
Operation is identical to Error screen. See Warning
Register 9.8.3 and interface description D184B093U31,
page 47.
Displays the communication software version.
i
Important:
Once the self-test function is initiated, the transmitter no
longer operates, i.e. the outputs no longer track the
flowrate.
Test the memory. Internal FRAM, ext. FRAM and calc.
Flash can be queried in order via ENTER function.
This menu does not appear for the PROFIBUS PA or
FOUNDATION fieldbus.
Test the pulse output. Enter a value directly in Hz.
This menu does not appear for the PROFIBUS PA or
FOUNDATION fieldbus.
Test the current output. Check the value displayed at the
connection terminals using a digital voltmeter or the
process instrumentation. Enter a value directly in mA.
No selection available if the HART device address is
anything other than 0.
This menu does not appear for the PROFIBUS PA or
FOUNDATION fieldbus.
Test the contact input. Before a signal is applied to the
contact input, the text "Off" is displayed in this menu.
Once a signal is applied, the text "On" is displayed (to
check via simulator, use the Totalizer Reset key).
Test the contact output. When you set the value to "On",
the contact output selected is displayed for the configured
contact logic under the item Switch Contacts - Contact
Output.
Test the display. Every point in the dot matrix is activated
one after another. Once all points are activated, a missing
point is easily detected.
Only for checking the transmitter using Simulator
55XC4000.
FSM4000
D184B140U02