12 Technical data
Pos: 48.2 /==== Wechsel ein- auf zweispaltig ==== @ 0\mod_1130421847171_3101.doc @ 3828
Wechsel ein-auf zweispaltig
Pos: 48.3 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Allgemein/Messgenauigkeit @ 3\mod_1157629438861_3101.doc @ 39739
12.1 Measuring accuracy
12.1.1 Reference conditionsper EN 29104
Fluid Temperatures
Ambient temp.
Supply power
Installation Conditions
Warm Up Phase
12.1.2 Maximum Measurement Error
Pulse output
−
DN 1 ... DN 2 (1/25 ... 1/12"):
± 1% of measured value, ± 0.001 Q
−
DN 3 ... DN 1000 (1/10 ... 40"):
Q > 0.05 Q
± 0.5% of measured value
maxDN
−
Q < 0.05 Q
± 0.00025 Q
maxDN
Q
= maximum flowrate for the flowmeter size 10 m/s
maxDN
Analog Output Effects
Same as pulse output plus ±0.1% of rate ±0.01 mA
6,0 %
5,0 %
4,0 %
Y
3,0 %
2,0 %
1,0 %
0,0 %
0
2
4
0
6,5
13,1
Fig. 59
Y
Accuracy ± of rate
X
Flow velocity
Pos: 48.4 /======= Spaltenumbruch ======== @ 0\mod_1132937966324_3101.doc @ 3831
D184B140U02
20 °C (68 °F) ± 2 K
20 °C (68 °F) ± 2 K
Line voltage per type plate U
N
−
upstream > 10 x DN
straight section
−
downstream > 5 x DN
straight section
DN = Flowmeter sensor size
30 min
maxDN
maxDN
DN 1...2
(1/25...1/12")
DN 3...1000
(1/10...40")
6
8
10 m/s
19,7
26,2
32,8 ft/s
X
Pos: 48.5 /Überschriften/1.1/2-spaltig/M - O/Messwertaufnehmer @ 15\mod_1195024106281_3101.doc @ 140273
12.2 Flowmeter sensor
Pos: 48.6 /Überschriften/1.1.1/2-spaltig/Messwertaufnehmer SE41F @ 17\mod_1200930203802_3101.doc @ 153533
12.2.1 Flowmeter sensor SE41F
Pos: 48.7 /Technische Daten / Datenblatt/Durchfluss/FSM4000/Technische Daten/Messaufnehmer/Modell SE41F @ 3\mod_1157634558052_3101.doc @ 39949
Temperature graph
± 1%
Fig. 60:
Fluid temperature as function of ambient temperature
1
Ambient temperature °C
2
Fluid temperature °C
3
Standard flange (steel): Hard/soft rubber max. 90/60 °C
(194 ... 140 °F)
4
with stainless steel flange
5
Standard flange (steel): PTFE/PFA/ETFE max.130 °C (266°F)
6
High temperature: Thick PTFE / PFA max. 180 °C (356 °F)
Max. allowable cleaning temperature PTFE-, PFA-design
CIP-Cleaning
Steam cleaning PTFE, PFA 150 °C (302 °F)
Liquid cleaning
If the ambient temperature is > 25 °C, then the difference must be
subtracted from the max. cleaning temperature.
- Δ °C. Δ °C = (T
T
max
Weight
Refer to dimensional drawings
G00456
Min. allowable pressure as factor of fluid temperature
Liner
Hard rubber
Soft rubber
PTFE
Thick PTFE
High
temperature
Design
PFA
ETFE
FSM4000
Technical data
1
3
4
Liner
T
max
Flowmeter
sensor
PTFE, PFA 140 °C (284 °F)
- 25 °C)
amb.
Nominal size
P
at
Operation
DN
mbar abs
15 ... 1000
0
(1/2 ... 40")
50 ... 1000
0
(2 ... 40")
10 ... 600
270
(3/8 ... 24")
400
500
25 ... 80
0
(1 ... 3")
100 ... 250
67
(4 ... 10")
300 (12")
27
3 ... 200
0
(1/10 ... 8")
0
25 ... 1000
100
(1 ... 40")
2
5
6
G00214
t
T
max
amb.
Min.
60
25 °C (77 °F)
60
25 °C (77 °F)
T
Operating
< 90 °C (194 °F)
< 60 °C (140 °F)
< 20 °C (68 °F)
< 100 °C (212 °F)
< 130°C (266 °F)
< 180 °C (356 °F)
< 180 °C (356 °F)
< 180 °C (356 °F)
< 130°C (266 °F)
< 180 °C (356 °F)
< 130 °C (266 °F)
111