8. Troubleshooting
Symptom
【LD ERR】
【FLOW ERR】
【BATTERY ERR】
【OVER ERR】
【STORAGE ERR】
【Alarm】
www.actoolsupply.com
Possible Cause
The laser may be damaged.
The flow channel may be interrupted.
The pump may be damaged.
Power capacity is low.
The instrument may be being used
outside the operating environment in the
specifications.
The remaining capacity of CF card is too
low.
Reading exceeds the Alarm setting value.
www.actoolsupply.com
Maintenance
Return the instrument to your distributor or to your
KANOMAX service center for repair.
Make sure that the inlet or outlet is not blocked.
Return the instrument to your distributor or to your
KANOMAX service center for repair.
If using a rechargeable battery, charge the battery.
If using AC line, the line may have problems.
Inside the optical system may be dirty.
Attach the zero filter to the inlet and perform aging.
Delete unnecessary data inside the CF card.
-