3. Display Description and Operation Procedure
[REPEAT MODE]
Configure SAMPLE T and CYCLES, and repeat measuring.
Example: If the setting is as follows in "CONFIG" menu (Refer to 3.6.4 REPEAT):
INTERVAL
SAMPLE T.
CYCLES
20 measurement cycles lasting 15 seconds each are taken.
Measurement Flow
Wait Time
Start
Measurement
[STAT MODE]
Configure SAMPLE T., CYCLES and SAMPLE POINT, and repeat measuring.
Measurement at each point is same as the one in Repeat Mode.
As for the measurement condition, you can configure the settings referring to 3.6.5 STAT
in 3.6 CONFIG.
[STANDARD MODE]
You can perform a measurement which complies with 7 measurement standards as follows:
-
ISO14644-1
-
ISO 14644-1 SEQUENTIAL SAMPLING
-
FEDERAL STANDARD (m)
-
FEDERAL STANDARD (ft)
-
BRITISH STANDARD
-
EC GMP
-
GB/T 16292-1996
Select and set one of the above measurement standards referring to STANDARD in page 35
in 3.9.2 OPTION.
As for the measurement condition, you can configure the settings referring to 3.6.6
STANDARD in 3.6 CONFIG.
www.actoolsupply.com
: N/A
: 00:00:15
: 20
Sampling Time
Sampling Time
(15sec)
(15sec)
st
nd
The 1
The 2
Measurement
Measurement
(International Standard) Refer to ISO14644-1.
(International Standard) Refer to ISO14644-1.
(The US Standard) Refer to FEDERAL STANDARD 209E.
(The US Standard)
(British Standard) Refer to BS-5295.
(European Standard) Refer to EC GMP.
(Chinese Standard) Refer to GB/T 16292-1996.