Emerson Rosemount 3051 Reference Manual page 149

Pressure transmitter with hart protocol
Hide thumbs Also See for Rosemount 3051:
Table of Contents

Advertisement

Reference Manual
00809-0100-4001, Rev JA
Pressure Upper Range Limit - Configurable Description
Standard
1
30 psi (2,1 bar)
2
150 psi (10,3 bar)
3
800 psi (55,2 bar)
4
4000 psi (275,8 bar)
5
10000 psi (689,5 bar)
Transmitter Output
Standard
A
4–20 mA with Digital Signal Based on HART Protocol
F
F
fieldbus Protocol
OUNDATION
(2)
W
Profibus PA Protocol
Expanded
M
Low-Power 1–5 Vdc with Digital Signal Based on HART Protocol
Process Connection Style
Standard
1
2B
/
–14 NPT Female
2
2C
G½ A DIN 16288 Male (Available in SST for Range 1–4 only)
Expanded
2F
Coned and Threaded, Compatible with Autoclave Type F-250-C (Includes Gland and Collar, Available in
SST for Range 5 only)
61
Non-threaded Instrument flange (Range 1-4 only)
Isolating Diaphragm
Standard
(3)
2
316L SST
(3)
3
Alloy C-276
Sensor Fill Fluid
Standard
1
Silicone
2
Inert (Fluorinert
Specifications and reference data
(1)
3051TG
®
FC-43)
Appendix A: Reference Data
3051TA
30 psia (2,1 bar)
150 psia (10,3 bar)
800 psia (55,2 bar)
4000 psia (275,8 bar)
10000 psia (689,5 bar)
Process Connection Wetted Parts
Material
316L SST
Alloy C-276
November 2012
Standard
Standard
Standard
Standard
Standard
139

Advertisement

Table of Contents
loading

Table of Contents