Running Time: 73 min
Preparation before using
1. Add 1.0 ml RNase Free Water to the Carrier RNA tube and mix by vortexing. Store prepared Carrier RNA (1mg/ml) at -20°C.
2. Add 1.1ml PK Storage Buffer to the Proteinase K tube and mix by vortexing. Store prepared Proteinase K (10mg/ml) at -20°C.
Protocol
1. Pipet 10 μl Carrier RNA(1mg/ml) and 20 μl proteinase K(10mg/ml) into the MagCore Sample Tubes(provided).
2. Add 1200 μl of serum, plasma, or cell-free body fluids into the prepared Sample Tube.
3. Place the prepared Sample Tube into well 4 of T-rack.
4. Put the Elution Tube into well 1 of T-Rack and the Tip Plus Holder Set into well 2 of T-Rack.
5. Run Code.211 program at MagCore®.
(sample volume :1200 μl)
MagCore®
24