4.2.4.3 Timer 1

Purpose:
Type:
Read Command:
Response:
Write Command:
Response:
Data Format:
Data Range:
4.2.4.4 Timer 2
Purpose:
Type:
Read Command:
Response:
Write Command:
Response:
Data Format:
Data Range:
4-12
To set Timer 1 or to read the current Timer 1 setting. Timer 1 is used when the camera is
operating in ExSync programmable mode or in free-run mode. See Section 3.2 for details.
This is a read or write command.
Cmd-ID
0xA6
Cmd-ID
0xA6
Cmd-ID
0xA6
None
Byte 1
Byte 2
Byte 3
Timer 1 settings can range from 6 µs to 16777215 µs
(0x000006 to 0xFFFFFF).
To set Timer 2 or read the current Timer 2 setting. Timer 2 is used when the camera is
operating in free-run mode. See Section 3.2 for details.
This is a read or write command.
Cmd-ID
0xA7
Cmd-ID
0xA7
Cmd-ID
0xA7
None
Byte 1
Byte 2
Byte 3
Timer 2 settings can range from 1 µs to 35 µs (0x000001 to 0xFFFFFF).
DRAFT
R/W-Flag
Data Length
1
R/W-Flag
Data Length
0
R/W-Flag
Data Length
0
Low byte of the Timer 1 setting
Mid byte of the Timer 1 setting
High byte of the Timer 1 setting
R/W-Flag
Data Length
1
R/W-Flag
Data Length
0
R/W-Flag
Data Length
0
Low byte of the Timer 2 setting
Mid byte of the Timer 2 setting
High byte of the Timer 2 setting
Data
3
-
Data
3
3 Bytes
Data
3
3 Bytes
Data
3
-
Data
3
3 Bytes
Data
3
3 Bytes